Back to Products
Ador Digatron MCT HD cell cycler cabinet, three-quarter side view

Cell & Battery Test Solutions · Lead Test & Formation · Laboratory · Cycler · Cell Tester

MCT HD

This cycler is the right choice for automated tests on battery cells in the current range from 2 amps to 40 amps. Thanks to its high packing density — unique on the market — it offers up to 384 test circuits in just one cabinet, making it perfect for laboratories where high quantities of cells need to be tested.

  • Precise Test Results

    Four selectable current ranges with hardware and software perfectly synchronized for precise measurement results.

  • High Energy Efficiency

    Switch-mode amplifiers with up to 95% efficiency. Regenerative DC links feed energy back to other test circuits.

  • Maximum Scalability

    Scales from a few circuits with one plug-in unit to several thousand, growing easily with your demands.

  • Incredibly Space-Saving

    A packing density unique on the market takes up little valuable laboratory space while ensuring high test capacities.

  • Parallel Connection of Circuits

    Two or four power outputs can be connected in parallel to increase the current accordingly.

Technical Data

Voltage rangeCharge: -0.1 V up to 6 V · Discharge: 0.7 V up to 6 V
Current range of circuits2 A up to 40 A
Paralleling of circuitsYes
Control accuracy U DC< ± 0.025 % fs · ± 0.05 % fs ≥ 10 A
Control accuracy I DC< ± 0.025 % fs · ± 0.05 % fs ≥ 10 A
Regenerative DC linkYes
Mains feedbackOptional
Circuits per power module4 up to 64
Max. circuits per cabinet384
ScalableYes

* additional values on request

Suitable for the Following Tests

CapacityCyclic tests to determine agingEfficiencyCustomized test profiles